Request for authorisation of a clinical investigation
2022.08.17
Clinical investigations of medical devices that are conducted as part of the clinical evaluation for conformity assessment purposes and where the investigational device either may not yet bear a CE mark or its use in the clinical investigation is outside its intended purpose require approval by the competent federal higher authority.
Legal Basis
The obligation to obtain an authorisation from the competent authority in the respective Member State of the European Union arises from Article 62(4) point a of Regulation (EU) 2017/745 (MDR):
„(a) the clinical investigation is the subject of an authorisation by the Member State(s) in which the clinical investigation is to be conducted, in accordance with this Regulation, unless otherwise stated;“
Application and authorisation procedure
According to Article 70 MDR, an application for authorisation of a clinical investigation must be submitted via the electronic system pursuant to Article 73 MDR. As long as the central European database EUDAMED does not provide the corresponding functionality, the application is submitted via the German Medical Devices Information and Database System (DMIDS) of the Federal Institute for Drugs and Medical Devices (BfArM). With the start of application of the MDR, a sequential approval procedure was introduced in Germany: The application must first be forwarded to the competent ethics committee and can only be transmitted to the competent federal higher authority after its approval (all via DMIDS).
The application process, including the procedure and deadlines at the authority, is described in more detail in Article 70 MDR and can be split into a validation phase and a substantive review phase. During the validation of an application, the authority ensures that the application falls within the scope of the MDR and that all required documents have been provided by the sponsor (see Article 70(2)-(5) MDR). This validation phase is identical for all applications submitted to the federal higher authority for authorisation of a clinical investigation according to MDR. The subsequent substantive review of the application differs according to the regulatory classification of the investigational device in accordance with Article 70(7) MDR in conjunction with Section 31 of the Medical Device Law Implementation Act (“Medizinprodukterecht-Durchführungsgesetz”, MPDG).
It should be noted in particular that Germany has made use of the option to define deviating national regulations:
„§ 31 Commencement of a Clinical Investigation
(1) A clinical investigation of devices classified as class I in accordance with the classification rules laid down in Chapter III of Annex VIII to Regulation (EU) 2017/745 or of non-invasive devices classified as class IIa in accordance with the classification rules laid down in Chapter III of Annex VIII to Regulation (EU) 2017/745 may only be commenced after
1. the competent federal higher authority has not objected within ten days of the validation date in accordance with Article 70(5) of Regulation (EU) 2017/745; and
2. the competent ethics committee pursuant to Article 33(1) has issued a favourable opinion in this regard.
(2) A clinical investigation of devices other than those referred to in paragraph 1 may only be commenced if
1. he competent federal higher authority has granted approval for this purpose, and
2. the competent ethics committee pursuant to Section 33 paragraph 1 has issued a favourable opinion in this regard.“
The following should be noted:
In Germany, in deviation from the requirements of the MDR, the full authorisation procedure according to Article 70(7) point b MDR and not the abridged authorisation procedure according to Article 70(7) point a MDR must be chosen for investigational devices that are assigned to class IIb according to the classification rules of Annex VIII Chapter III of Regulation (EU) 2017/745. This therefore also applies to non-invasive investigational devices of risk class IIb.
Furthermore, contrary to the wording in the MDR, in the case of investigational devices of risk class I and non-invasive IIa (abridged authorisation procedure), the clinical investigation may not be started immediately after the date of validation. According to Section 39(3) MPDG, the competent federal higher authority must, according to Section 31(1) MPDG (see above, abbreviated application procedure), check within 10 days from the date of validation "...whether the classification rules of Annex VIII of Regulation (EU) 2017/745 have been applied correctly."