Product group Medical instruments for use in humans
Reference 02625/07

Recall of the products TOM PLUG FOR 01-15400/401 (lots see annex) from Stryker, because the plug´s dimension did not meet the specified vaules. This caused friction with the cutting cylinder during use. The potential risk is that metal shavings come off the plug as a result of the friction between plug and cutting cylinder.

Recall of the products TOM PLUG FOR 01-15400/401 from Stryker , Download_VeroeffentlichtAm_EN PDF, 75KB, File does not meet accessibility standards