Field Safety Notice for Path Finder and InCompass T-Handle Driver from Abbott Spine Inc.

Product group Medical instruments for use in humans
Reference 04524/08

The manufacturer Abbott Spine Inc. initiated a field safety corrective action regarding Path Finder and InCompass T-Handle Driver (Models 2151-4, 2151-5, 2151-6). An analysis had determined that the instruments were not being assembled properly after cleaning due to insuffifient cleaning instructions.

Field Safety Notice for Path Finder and InCompass T-Handle Driver from Abbott Spine Inc. , Download_VeroeffentlichtAm_EN PDF, 808KB, File does not meet accessibility standards