BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Smiths Medical MD reminds in a Safety Alert about problems with the CADD-Legacy® 1 or CADD-Legacy® PLUS Ambulatory Infusion Pump

Product group Injections / Infusions / Transfusions / Dialysis
Reference 00416/06

The company Smiths Medical MD reminds customers, distributors and affiliates in a Safety Alert of different units for the infusion rate when programming CADD-Legacy® 1 or CADD-Legacy® PLUS Ambulatory Infusion Pumps. The first mentioned pump is designed to infuse medication in ml/24h, the latter in ml/h.

Smiths Medical MD reminds in a Safety Alert about problems with the CADD-Legacy® 1 or CADD-Legacy® PLUS Ambulatory Infusion Pump , Download_VeroeffentlichtAm_EN PDF, 153KB, File does not meet accessibility standards