Information about B|Braun Space Perfuso from B. Braun Melsungen

Product group Injections / Infusions / Transfusions / Dialysis
Reference 02686/06

B. Braun Melsungen has issued a customers information concerning the B|Braun Space Perfusor- and Infusomat with prime function. After the prime process is finished the occlusion pressure does not automatically fall back to the manual pressure setting but remains at the maximum.