Safety-relevant recommendation for Applix Smart / Applix Vision, Fresenius Kabi

Product group Injections / Infusions / Transfusions / Dialysis
Reference 01794/08, 01309/08

The company Fresenius Kabi notified customers of the Enteral Feeding Pumps Applix Smart / Applix Vision that the pumps will deliver an initial volume of 4 ml (resp. 2 ml) directly after the start. These deliveries must be evaluated particularly on infants and neonates.The same customer letter contains a reference to a possible display failure where not all segments of the figure are shown correctly. Users are recommended to pay attention to the display during the self test.

Safety-relevant recommendation for Applix Smart / Applix Vision, Fresenius Kabi , Download_VeroeffentlichtAm_EN PDF, 120KB, File does not meet accessibility standards