BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice of battery caps in insulin pumps

Product group Injections / Infusions / Transfusions / Dialysis
Reference 04457/08

Med Trust Handelsgesellschaft mbH exchanges battery caps in insulin pumps manifactured by Animas Corporation.

Field Safety Notice of battery caps in insulin pumps , Download_VeroeffentlichtAm_EN PDF, 28KB, File does not meet accessibility standards

Use of cookies

Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)

OK