BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information of Luer connector of specific A/V and S/N haemo­dialysis tubing systems for the 5008

Product group Injections / Infusions / Transfusions / Dialysis
Reference 04497/08

The manufacturer Fresenius points out that users might insert the Luer connector of specific A/V and S/N haemo­dialysis tubing systems for the 5008 dialysis machine gently and avoid screwing the Luer-Lock tightly. Otherwise, the pro- or post dilution port may brake during dis­connection, possibly leading to leakage of extracorporeal blood volume.

Information of Luer connector of specific A/V and S/N haemo­dialysis tubing systems for the 5008 , Download_VeroeffentlichtAm_EN PDF, 2MB, File does not meet accessibility standards