Safety Notice for 5008 tubing systems, Fresenius

Product group Injections / Infusions / Transfusions / Dialysis
Reference 04497/08

Fresenius Medical Care informs per Field Safety Notice about the safe handling of the tubing systems for the hemodialysis systems 5008. The pre- or post-dilution port of the tubing system may break if the male Luer-Lock connector is connected too forceful and firm.