BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of Dade Behring for Thromborel S

Product group In-vitro diagnostics
Reference 03058/04

Corrective action of Dade Behring for Thromborel S (lot number 505696B, 505696C, 505697B, 505697C, 505698B, 505699B, 505699C, 505699D, 505701B and 505701C). Certain ISI values provided from the manufacturer for Sysmex CA 500 series analysers are erroneous and may cause falsely increased INR values. Customers were informed about the problem and advised to use the corrected values.

Corrective action of Dade Behring for Thromborel S , Download_VeroeffentlichtAm_EN PDF, 68KB,