Corrective action and recall of BioMerieux for the BacT/Alert Classic 120 System and the BacT/Alert 240 Classic System

Product group In-vitro diagnostics
Reference 01581/05

Corrective action and recall of BioMerieux for the BacT/Alert Classic 120 System and the BacT/Alert 240 Classic System because of the risk of false negative test results if the system is used in combination with generic bar codes. The customers were informed about the problem and how to avoid it. A new software version will be developed.

Corrective action and recall of BioMerieux for the BacT/Alert Classic 120 System and the BacT/Alert 240 Classic System , Download_VeroeffentlichtAm_EN PDF, 239KB,