Recall for reagent STA-Neoplastin Plus, Diagnostica Stago
2008.09.08
Product group In-vitro diagnostics
Reference 02945/08
Recall of Diagnostica Stago for reagent STA-Neoplastin Plus (Ref. 00142, Lot 684816) because of an inhomogenity of this batch giving prolonged prothrombin times and leading to QC values outside of the ranges. The customers were requested to stop the use.
Recall for reagent STA-Neoplastin Plus, Diagnostica Stago , Download_VeroeffentlichtAm_EN