Safety Notice for Bicarbonate FS, DiaSys Diagnostic Systems

Product group In-vitro diagnostics
Reference 03887/09

DiaSys Diagnostic Systems has issued an information about a reduced stability of the reagent Bicarbonate FS when used in an open vial. After 3 days with opened vial the linearity of the test was affected leading to falsely low measuring values.

Safety Notice for Bicarbonate FS, DiaSys Diagnostic Systems , Download_VeroeffentlichtAm_EN PDF, 43KB, File does not meet accessibility standards