Corrective Action for Analyzers ABL505/555, EML105, ABL600, ABL700 Series, ABL800FLEX and ABL800BASIC, Radiometer

Product group In-vitro diagnostics
Reference 00169/10

Corrective Action of Radiometer for Analyzers ABL505/555, EML105, ABL600, ABL700 Series, ABL800FLEX and ABL800BASIC because the drug Irenat (perchlorate) interferes with ionized calcium resulting in a negative bias. The customers were informed about the problem and how to avoid it.

Corrective Action for Analyzers ABL505/555, EML105, ABL600, ABL700 Series, ABL800FLEX and ABL800BASIC, Radiometer , Download_VeroeffentlichtAm_EN PDF, 50KB, File does not meet accessibility standards