Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2010.02.19
Product group In-vitro diagnostics
Reference 00169/10
Corrective Action of Radiometer for Analyzers ABL505/555, EML105, ABL600, ABL700 Series, ABL800FLEX and ABL800BASIC because the drug Irenat (perchlorate) interferes with ionized calcium resulting in a negative bias. The customers were informed about the problem and how to avoid it.