Recall for the Complement reagent, Siemens

Product group In-vitro diagnostics
Reference 03459/10

Recall of Siemens for the Complement reagent (OWZD11, Lot no. 542865) because falsely increased or decreased values may be observed due to a high CV of duplicate determinations leading to invalid calibration curves and high variability in control recovery on the BCS/BCS XP Systems. Customers were advised not to use the affected lot on the BCS/BCS XP Systems.

Recall for the Complement reagent, Siemens , Download_VeroeffentlichtAm_EN PDF, 50KB, File does not meet accessibility standards