Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2010.10.25
Product group In-vitro diagnostics
Reference 03543/10
Recall of Roche for Calcium reagent because of imprecision and erroneous patient results for Calcium due to precipitations in R2. The customers were informed about the problem and requested to stop using the concerned lots.
Recall for Calcium reagent, Roche , Download_VeroeffentlichtAm_EN