Recall/Corrective Action for Coulter HmX and HmX AL Hematology Analyzers and for Coulter MAXM and MAXM AL Hematology Analyzers, Beckman Coulter

Product group In-vitro diagnostics
Reference 00016/11

Recall/Corrective Action of Beckman Coulter for Coulter HmX and HmX AL Hematology Analyzers and for Coulter MAXM and MAXM AL Hematology Analyzers (all Software Revisions) because the tilde (~) character is omitted. There is a potential for Specimen or Patient misidentification with serious consequences. The customers were informed about the problem and requested to do not use the tilde (~) character in Specimen or Patient identifiers.

Recall/Corrective Action for Coulter HmX and HmX AL Hematology Analyzers and for Coulter MAXM and MAXM AL Hematology Analyzers, Beckman Coulter , Download_VeroeffentlichtAm_EN PDF, 48KB, File does not meet accessibility standards