BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Cardiac Science recalls the Powerheart AED model 9200RD und 9212RD

Product group Medical electronics / Electromedical devices
Reference 02513/04

Cardiac Science has initiated a recall of the Powerheart AED model 9200RD und 9212RD that were shipped to customers in the late 2002 and early 2003. A marginal capacitor component may cause a longer than specified charge time and lower than specified energie output.

Cardiac Science recalls the Powerheart AED model 9200RD und 9212RD , Download_VeroeffentlichtAm_EN PDF, 88KB,