Corrective action for AED 10 defibrillators, Welch Allyn

Product group Medical electronics / Electromedical devices
Reference 00487/07

Welch Allyn Ltd. has issued an information concerning a corrective action of the AED 10 defibrillators. There is the possibility of an early low battery alarm. But after appearance of this indication the device is capable of delivering at least 15 discharges at 360 joules of shocks of therapy. Therefore the customers are recommended to still use the defibrillators until a replacement battery pack is obtained.

Corrective action for AED 10 defibrillators, Welch Allyn , Download_VeroeffentlichtAm_EN PDF, 57KB, File does not meet accessibility standards