Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2007.08.08
Product group Medical electronics / Electromedical devices
Reference 02716/07
The Metrax GmbH has issued an information concerning a problem of the Primedic XD1, XD3, XD10, XD30, XD100, XD300, XD110 and XD330 defibrillators of a defined serial number range. The affected devices can stop due to PaddleLogicError 135 right after switching on. The failure can be fixed by disconnecting the device from the main power supply and a restart in battery mode.