BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Switched Reusable Internal Defibrillation Paddles, Philips

Product group Medical electronics / Electromedical devices
Reference 00621/10

Philips Healthcare has issued a Field Safety Notice concerning the Switched Reusable Internal Defibrillation Paddles M4741A, M4742A, M4743A and M4744A used as accessories for the external defibrillators HeartStart XL and HeartStart MRx. The shock switch on the handle in paddles manufactured from September 1, 2008 to December 30, 2008 may fail to actuate, so that delivery of defibrillation energy will be delayed or prevented. The affected paddles are replaced.

Safety Notice for Switched Reusable Internal Defibrillation Paddles, Philips , Download_VeroeffentlichtAm_EN PDF, 136KB, File does not meet accessibility standards

Use of cookies

Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)