Safety Notice for Corpuls3, GS Elektromedizinische Geräte G. Stemple GmbH

Product group Medical electronics / Electromedical devices
Reference 03567/10

GS Elektromedizinische Geräte G. Stemple GmbH has issued a Field Safety Notice concerning a corrective action of the external defibrillator Corpuls3. A defective part of a circuit board may lead to a malfunction of the infrared communication between the connected modules during the use as a compact device. The defective component of the affected devices will be exchanged.

Safety Notice for Corpuls3, GS Elektromedizinische Geräte G. Stemple GmbH , Download_VeroeffentlichtAm_EN PDF, 257KB, File does not meet accessibility standards