Recall for Neurotrac™MyoPlus2, Verity Medical Ltd.

Product group Medical electronics / Electromedical devices
Reference 02687/11

The company Verity Medical Ltd. notified of a recall of the physical therapy electrical stimulation system Neurotrac™MyoPlus2. Due to a software anomaly a divergence between reading on the LCD display and intensity is possible. A sudden surge of stimulation on one of the output channels could be experienced by users as unpleasant shock sensation.

Recall for Neurotrac™MyoPlus2, Verity Medical Ltd. , Download_VeroeffentlichtAm_EN PDF, 116KB, File does not meet accessibility standards