BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of the BIS Bilateral EEG (electroencephalography) Sensors, Covidien

Product group Medical electronics / Electromedical devices
Reference 04721/11

Covidien has issued a Field Safety Notice concerning a recall of the BIS Bilateral EEG (electroencephalography) Sensors. The reference and left eye electrode circuit traces of several lot numbers from part number 186-0212 are reversed. This failure will result in the inaccurate calculation and presentation of processed EEG information.

Recall of the BIS Bilateral EEG (electroencephalography) Sensors, Covidien , Download_VeroeffentlichtAm_EN PDF, 324KB, File does not meet accessibility standards