BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Urgent Safety Notice for the Samaritan PAD SAM 300/300P public access defibrillators, HeartSine Technologies Ltd.

Product group Medical electronics / Electromedical devices - electrotherapy
Reference 05995/12

HeartSine Technologies Ltd. has issued a Field Safety Notice concerning corrective actions of Samaritan PAD SAM 300/300P public access defibrillators.
1. Devices manufactured between August 2004 and December 2010 may turn theirselves on and of repeatedly so that the battery will become completely depleted.
2. Devices containing early software versions distributed until end December 2010 may issue a low battery warning even though sufficient battery capacity remains.

Urgent Safety Notice for the Samaritan PAD SAM 300/300P public access defibrillators, HeartSine Technologies Ltd. , Download_VeroeffentlichtAm_EN PDF, 2MB, File does not meet accessibility standards

Use of cookies

Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)

OK