Urgent Safety Notice for PROCARE Monitors B20 or B40, Software Version VSP-A_1.10 (or earlier), GE Healthcare

Product group Medical electronics / Electromedical devices - electrodiagnostics
Reference 06820/12

At Monitors B20 or B40 with Software Version VSP-A_1.10 (or earlier) the ‘V Tachy’ and ‘V Fib’ alarms are disabled in NEONATAL mode. Other modes and parameters are not affected.