BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Urgent Safety Notice for Patient Data Module (PDM) Version 2.0,GE Healthcare

Product group Medical electronics / Electromedical devices - electrodiagnostics
Reference 07981/12

When the certain monitoring systems Solar 8000i, Transport Pro, CARESCAPE Monitor B850, CARESCAPE Monitor B650 using a Patient Data Module are configured for auto-mode Non-Invasive Blood Pressure measurements, it is possible that the auto mode measurements may stop automatically cycling.

Urgent Safety Notice for Patient Data Module (PDM) Version 2.0,GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 69KB, File does not meet accessibility standards