BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of the Rexpol-Einsätze Standard from Plus Orthopedics

Product group Non-active implants
Reference 01497/07

Recall of the products HI Rexpol-Einsätze Standard (article numbers and lots see annex) from Plus Orthopedics AG. The PE-plug which should be inserted in the Rexpol-inset during hip implantation does not fit in the hole of the Ti-cup. This might cause a reduction of the rotation securing of the Rexpol-inset in the Ti-cup. The firm reported, that the security of the implanted prostheses is warranted without after­care measures.

Advisory Notice is only available in German language.

Recall of the Rexpol-Einsätze Standard from Plus Orthopedics , Download_VeroeffentlichtAm_EN PDF, 79KB, File does not meet accessibility standards