Safety Notice for VISIOL® pre-filled syringes, TRB Chemedica
2009.04.08
Product group Non-active implants
Reference 00292/09
TRB Chemedica amends the instructions for use of the VISIOL® pre-filled syringes. It is possible that the fingergrip can unintentionally slip off when applying a horizontal force if the syringe is not held in a correct manner as described in the instructions for use leaflet.
Safety Notice for VISIOL® pre-filled syringes, TRB Chemedica , Download_VeroeffentlichtAm_EN