BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Lot recall of Cannulated Screw 3.5, aap Implantate

Product group Non-active implants - bone surgery
Reference 06423/13

Cannulated Screw 3.5: The company aap Implantate has informed about a lot recall of devices. The action was initiated because of a wrong length designation on the label of screws (44 instead of 46 mm and vice versa).

Lot recall of Cannulated Screw 3.5, aap Implantate , Download_VeroeffentlichtAm_EN PDF, 304KB, File does not meet accessibility standards