BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Sola 300 examination lights, Dräger Medical

Product group Surgical equipment/ Anaesthesia
Reference 01604/10

Dräger Medical AG & Co KG has issued a Field Safety Notice concerning a corrective action on spring-loaded arms in Sola 300 examination lights installed between 2001 and 2003. Spring arms with a plastic joint, which can be identified by its bent extension arm, carry under certain circumstances an increased risk of the joint breaking suddenly and unexpectedly. The affected spring-loaded arms will be exchanged for the latest spring-loaded arm with aluminium alloy casting joint.

Safety Notice for Sola 300 examination lights, Dräger Medical , Download_VeroeffentlichtAm_EN PDF, 241KB, File does not meet accessibility standards

Use of cookies

Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)