BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of surgical lighting systems iLED and TruLight, Trumpf Medizin Systeme GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia
Reference 05933/11

Trumpf Medizin Systeme GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the surgical lighting systems iLED and TruLight. Due to a tight handle socket the Sterilisable Handle ALC, material number 1583966, may not lock in the catch provided and as consequence of this it may fall off into the surgical field. All handles of this type will be replaced with new handles made with an increased tolerance.