Safety Notice for the Legendair™ and Supportair™ Ventilators, COVIDIEN

Product group Surgical equipment/ Anaesthesia - anaesthesia and medical gas supply
Reference 02313/12

Covidien has issued a Field Safety Notice concerning the possible failure of the ventilators. The defect of a capacitor may result in the inability of the ventilator to switch automatically from A/C mains power to internal battery operation when the ventilator is disconnected from A/C mains during use or when there is a failure of the A/C mains power.

Safety Notice for the Legendair™ and Supportair™ Ventilators, COVIDIEN , Download_VeroeffentlichtAm_EN PDF, 2MB, File does not meet accessibility standards