Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2013.10.21
Product group Surgical equipment/ Anaesthesia - anaesthesia and medical gas supply
Reference 05431/13
CareFusion has issued a Field Safety Notice concerning AVEA ventilators. In cases where a patient is ventilated at an higher altitude the ventilator may under report the delivery of tidal volume due to an error in the barometric pressure compensation.
Field Safety Notice concerning AVEA ventilators, CareFusion , Download_VeroeffentlichtAm_EN