BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning AVEA ventilators, CareFusion

Product group Surgical equipment/ Anaesthesia - anaesthesia and medical gas supply
Reference 05431/13

CareFusion has issued a Field Safety Notice concerning AVEA ventilators. In cases where a patient is ventilated at an higher altitude the ventilator may under report the delivery of tidal volume due to an error in the barometric pressure compensation.

Field Safety Notice concerning AVEA ventilators, CareFusion , Download_VeroeffentlichtAm_EN PDF, 74KB, File does not meet accessibility standards