Recall for the Rüsch Bronchopart® bronchial tubes, Teleflex Medical

Product group Surgical equipment/ Anaesthesia - anaesthesia and medical gas supply
Reference 00865/14

Teleflex Medical recalls several product and lot numbers of the Rüsch Bronchopart® bronchial tubes because the tube-cuff may inflate only to one side. Consequently there is the potential for oxygen de-saturation and loss of volatile anaesthetics, requiring re-intubation of the patient.

Recall for the Rüsch Bronchopart® bronchial tubes, Teleflex Medical , Download_VeroeffentlichtAm_EN PDF, 547KB, File does not meet accessibility standards