BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of one lot of the contact lens care solution OptiStar 2

Product group ophthalmological technology
Reference 02958/04

The German distributor OptiStar Großhandels GmbH is conducting a voluntary recall of one lot (231B05; catalogue number OPTI2360) of the contact lens care solution OptiStar 2. The firm got knowledge from one complaint, where discomfort occurred after the application of the neutralization solution. The examination of the affected lot revealed, that the solution "OPTiStar 2" from the among mentioned lot does not neutralize the disinfection solution "OptiStar 1" sufficiently.

Advisory notice is only available in German language.

Recall of one lot of the contact lens care solution OptiStar 2 , Download_VeroeffentlichtAm_EN PDF, 32KB,