Product group Radiological technology
Reference 04351/07

Philips Medical Systems has issued an information concerning a problem with the PACS System EasyVision RAD Release 4.2 V2. If any zoomed image is exported to a PACS then an electronic measurement on the PACS can give a wrong result because during the export process this image is sent with the new pixel matrix but the pixel size information in the image has not been changed. Philips will perform a software update version Easy Vision RAD R4.2 V2 LEVEL 7 SP 4 on all systems.

Information for PACS System EasyVision RAD Release 4.2 V2, Philips , Download_VeroeffentlichtAm_EN PDF, 449KB, File does not meet accessibility standards