Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2008.04.02
Product group Radiological technology
Reference 00948/08
Philips Medizin Systeme GmbH issued a Safety Notice about memory artefacts on patient images concerning Essenta DR X-ray systems with software version 1.0.2. Bright artefacts may appear on a patient image from a previous exposure, particularly when the detector receives direct radiation due to inappropriate collimation. A software upgrade to version 1.0.3 is planned.
Information for Essenta DR X-ray systems, Philips , Download_VeroeffentlichtAm_EN