Product group Radiological technology
Reference 02844/08

GE Healthcare has issued a customer letter concerning 3100 and 4100 cardiovascular x-ray systems. The displayed dose data may be underestimated by a minimum of 50% if the system is not rebooted after 6 days and 5 hours of continious operation. A software upgrade is planned.

Information for Innova 3100 and 4100, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 49KB, File does not meet accessibility standards