BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning Mammographie X-ray system Diamond

Product group Radiological technology
Reference 04803/08

GE Healthcare has issued a Field Safety Notice concerning Mammographie X-ray system Diamond with serial numbers within the range of 20566-20725. The manufacturer has inspected all systems because fastening screws may not have been tightened properly. The C-Arm could have fallen down.

Field Safety Notice concerning Mammographie X-ray system Diamond , Download_VeroeffentlichtAm_EN PDF, 248KB, File does not meet accessibility standards