BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice concerning use of Brainlab Scull Reference Base in combination with KLS Martin Cross-Drive screws, Brainlab AG

Product group Radiological technology
Reference 03354/11

Brainlab AG has issued a field safety notice concerning use of Brainlab Scull Reference Base in combination with KLS Martin Cross-Drive screws. Applying too much torque or tensile force to the head of the screw during insertion and fixation of the Skull Reference Base may cause the screw head to deform or to shear off.

Safety Notice concerning use of Brainlab Scull Reference Base in combination with KLS Martin Cross-Drive screws, Brainlab AG , Download_VeroeffentlichtAm_EN PDF, 81KB, File does not meet accessibility standards

Use of cookies

Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)