Safety Notice concerning MammoDiagnost DR X-ray systems, Philips

Product group Radiological technology
Reference 05624/11

Philips has issued a Field Safety Notice concerning MammoDiagnost DR X-ray systems since metal gantry column covers are not securely electrically grounded. Field Service Engineers will attach a grounding cable to the covers of the column. This corrective action applies to all systems with production date before May 2011.