Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2011.12.14
Product group Radiological technology
Reference 05624/11
Philips has issued a Field Safety Notice concerning MammoDiagnost DR X-ray systems since metal gantry column covers are not securely electrically grounded. Field Service Engineers will attach a grounding cable to the covers of the column. This corrective action applies to all systems with production date before May 2011.
Safety Notice concerning MammoDiagnost DR X-ray systems, Philips , Download_VeroeffentlichtAm_EN