Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2012.09.03
Product group Radiological technology - radiological equipment for vascular diagnostics
Reference 05192/12
GE Healthcare has issued a Field Safety Notice concerning a software update required for Innova 2121IQ and Innova 3131IQ X-ray systems. When a LIH (last image hold) functionality is requested and a short fluoroscopy exposure (stopping fluoroscopy exposure within 2 seconds) is started without waiting for the display of the LIH (last image hold) in review monitor, there is a potential for loss of imaging which requires a restart of the system.