BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Software Update required for Innova 2121IQ and Innova 3131IQ X-ray systems, GE Healthcare

Product group Radiological technology - radiological equipment for vascular diagnostics
Reference 05192/12

GE Healthcare has issued a Field Safety Notice concerning a software update required for Innova 2121IQ and Innova 3131IQ X-ray systems. When a LIH (last image hold) functionality is requested and a short fluoroscopy exposure (stopping fluoroscopy exposure within 2 seconds) is started without waiting for the display of the LIH (last image hold) in review monitor, there is a potential for loss of imaging which requires a restart of the system.

Software Update required for Innova 2121IQ and Innova 3131IQ X-ray systems, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 61KB, File does not meet accessibility standards