Anomaly in Eclipse Photon Single Pencil Beam Convolution from Varian
2005.11.22
Product group Radiotherapy / Radiation protection
Reference 02653/05
Varian informs about an anomaly in Eclipse Photon Single Pencil Beam Convolution (PBC) version 7.2.35. Truncated dose distributions for tangential fields may occur if the X1 or Y1 jaws are in an over-travel position.
Anomaly in Eclipse Photon Single Pencil Beam Convolution from Varian , Download_VeroeffentlichtAm_EN