BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Oncentra Visir, Nucletron B.V.

Product group Radiotherapy / Radiation protection
Reference 00160/10

Nucletron B.V. has issued a Field Safety Notice concerning a problem regarding the use of Oncentra Visir of the versions Oncentra® Record/Visir v2.0 und v2.1 and their service packs. The consequences of this error is that beams that have reached the planned number of fractions may still be treated, resulting in too many fractions delivered and an overdose to the patient.

Safety Notice for Oncentra Visir, Nucletron B.V. , Download_VeroeffentlichtAm_EN PDF, 122KB, File does not meet accessibility standards