Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2011.08.02
Product group Radiotherapy / Radiation protection
Reference 02414/11
Nucletron BV has issued a notice concerning a failure of the door interlock during the Source Alignment Procedure. The software in the upcoming TCC 2.1 release will be adapted in such a way that the Door Interlock must be closed before the source alignment procedure can be started.
Safety Notice for Flexitron HDR Afterloader, Nucletron , Download_VeroeffentlichtAm_EN