Safety Notice for the Clinac, Trilogy, Novalis Tx and Unique jaw carriers, Varian medical systems

Product group Radiotherapy / Radiation protection - radiation facilites with accelerators for teletherapy
Reference 05987/12

Varian medical systems has issued a Field Safety Notice concerning an anomaly that has been identified with the Clinac, Trilogy, Novalis Tx and Unique jaw carriers. The jaw carriers function to couple the X-jaws to their respective drive screw mechanisms used for controlling field size. Some jaw carriers were manufactured with the incorrect metal alloy, which can potentially cause the jaw carrier to crack.

Safety Notice for the Clinac, Trilogy, Novalis Tx and Unique jaw carriers, Varian medical systems , Download_VeroeffentlichtAm_EN PDF, 358KB, File does not meet accessibility standards