Recall of several Kimal Procedure Packs containing BD Plastipak 50 ml Luer Lock Syringes, Kimal Plc

Product group Dressings / Pads - others
Reference 03308/13

The manufacturer Kimal Plc informs about the recall of several Kimal Procedure Packs containing 50 ml Plastipak Luer Lock Syringes with old design stopper of the manufacturer Becton Dickinson. If syringes with old design stopper are used in syringe pumps that have been adjusted for the new stopper, there may be a delay in initiating an occlusion alarm.

Recall of several Kimal Procedure Packs containing BD Plastipak 50 ml Luer Lock Syringes, Kimal Plc , Download_VeroeffentlichtAm_EN PDF, 100KB, File does not meet accessibility standards