Use of cookies
Cookies help us to provide our services. By using our website you agree that we can use cookies. Read more about our Privacy Policy and visit the following link: Privacy Policy (only in German)
2013.05.22
Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 02470/13
Memmert GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the heating ovens Memmert UF450plus, UF750plus and IF750plus. Due to an error in the assembly a lack of firmness of the castor supports may occur. There is the risk that the directional castors could tip sideways. The castor supports of the affected devices must be reinforced according to enclosed Refitting Instructions.