BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning heating ovens Memmert UF450plus, UF750plus and IF750plus, Memmert GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 02470/13

Memmert GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the heating ovens Memmert UF450plus, UF750plus and IF750plus. Due to an error in the assembly a lack of firmness of the castor supports may occur. There is the risk that the directional castors could tip sideways. The castor supports of the affected devices must be reinforced according to enclosed Refitting Instructions.

Field Safety Notice concerning heating ovens Memmert UF450plus, UF750plus and IF750plus, Memmert GmbH & Co. KG , Download_VeroeffentlichtAm_EN PDF, 964KB, File does not meet accessibility standards